2022
HIOKI offers advanced quality test solutions for electrode Slurry.
-Three parameters indicate the mixture condition of the slurry based on impedance
-Hardware & Software ideal for slurry impedance measurement
-Data & Analysis tool for various types of slurry
Unique HIOKI Slurry Analysis technology provides quantified mixing quality of electrode slurry to approach your ideal Battery. It enables you to optimize the slurry process and helps detect quality deviations in future battery cell manufacturing upstream processes. Ultimately, this will lead to the stabilization of cell quality.
Slurry Analytical System Capabilities
Hioki's Slurry Analytical System measures slurry impedance and analyzes the information. The measurement data lets you assess the slurry’s impedance, while the analysis can aid in improving the slurry's mixture conditions.
Key Features
- A proprietary Hioki algorithm analyzes impedance measured values for LiB electrode slurries
- Easily measure impedance of slurry
- Analysis Results “DCR, Rratio, Uniformity†indicate electron conductivity of Slurry
- Cloud-based analysis software ensures you always have access to the latest version
- Able to choose license plan, fit the right solution for your needs
Chuyên cung cấp thiết bị đo lường, thử nghiệm, dụng cụ cầm tay,...
Analytical Software SA2632
Measurement Software SA2633
Electrode Cell SA9001
Test Fixture SA9002
Measurement conditions*
*If using an instrument other than the IM3536 or IM3536-01- Use the Electrode Cell SA9001. The analytical algorithm assumes use of the SA9001. - Check whether the Test Fixture SA9002 can be connected to the instrument. - Acquire data under the measurement conditions listed below. - Prepare a CSV file to send to the system.
The number of data required depends on the measurement frequency range.
(Reference value)
1 mHz to 200 kHz: 700 points
1 mHz to 8 MHz: 1000 points
4 Hz to 8 MHz: 500 points
4 Hz to 300 MHz: 1000 points
100 kHz to 300 MHz: 500 points
Available material categories
- The system uses the appropriate analytical algorithm to analyze the data based on the selected material category combination. - You may not be able to select some combinations, and some material categories may not be available. If you encounter this issue, perform the analysis using the default model. - There’s no need to specify material proportions. - In some cases, the system may not be able perform analysis.- Hioki plans to add material categories over time.