Rated output voltage | 50 V DC | 125 V DC | 250 V DC | 500 V DC | 1000 V DC |
Effective maximum indicated value | 100 MΩ | 250 MΩ | 500 MΩ | 2000 MΩ | 4000 MΩ |
Accuracy 1st effective measuring range MΩ | ±4% rdg. 0.200 to 10.00 | ±4% rdg. 0.200 to 25.0 | ±4% rdg. 0.200 to 50.0 | ±4% rdg. 0.200 to 500 | ±4% rdg. 0.200 to 1000 |
Lower limit resistance | 0.05 MΩ | 0.125 MΩ | 0.25 MΩ | 0.5 MΩ | 1 MΩ |
Overload protection | 600 V AC (10 s) | 600 V AC (10 s) | 600 V AC (10 s) | 600 V AC (10 s) | 1200 V DC (10 s) |
DC voltage range | 4.2 V (0.001 V resolution) to 1000 V (1 V resolution), 4 ranges, Accuracy: ±1.3% rdg. ±4 dgt., (Ranges in excess of 1000 V are not guaranteed for accuracy.) |
AC voltage range | 420 V (0.1 V resolution)/600 V (1 V resolution), 2 ranges, 50/60 Hz, Accuracy: ±2.3% rdg. ±8 dgt., (Ranges in excess of 600 V are not guaranteed for accuracy.) |
Display | Semi-transmissive FSTN LCD with back lighting, Backlight |
Response time | Insulation resistance range: 1 second, PVΩ function: 4 seconds (based on in-house tests) |
Other functions | Live circuit indicator, automatic electric discharge, automatic DC/AC detection, comparator, drop proof, auto power save |
Power supply | AA alkaline batteries (LR6) ×4, Continuous operating time: Approx. 20 hours (based on in-house tests) |
Dimensions and mass | 159 mm (6.26 in) W × 177 mm H (6.97 in) H × 53 mm (2.09 in) D, Approx. 600 g (21.2 oz) (including batteries, excluding test lead) |
Included accessories | TEST LEAD L9787 ×1, Neck strap ×1, Instruction manual ×1, AA alkaline batteries (LR6) ×4 |