Withstanding Voltage | AC 3700V, 1 minute (Between input terminal and outer case) |
Storage Temperature | -10℃~60℃, 80% RH max (Without condensation) |
Operating Temperature | 0℃ to 40℃, 80% RH max. (Without condensation) |
Power supply | 1.5V (AM-3, R6 or "AA" size)×6 |
Weight | Approx. 330g (Excluding batteries) |
Safety standard | IEC 61010-1 (2001) Installation categoryâ…¡, 600V phase to earth. |
Size | 170(W)×105(D)×54(H)mm |
Accessories | Line test lead:1 Earth test lead:1 Batteries:6 Test lead case:1 Belt:1 Instruction manual:1 |
Optional accessory | Test probe with remote switch |
E.M.C. standard | EN 61326 |
Function | Insulation resistance, AC voltage, Battery check |
Meter movement | 100μA, 870Ω, Taut band meter |
Constructional standard | In accordance with IEC1557-2 or JIS C1302 (1994) |
Overload protection | 120% of the highest nominal output voltage (10 sec.) |
Battery check | DC 6.3V~9.5V |
Low battery limit | DC 6.3V |
Temperature characteristics(0~40℃) | ±5% rdg of specified accuracy |
Accuracy(DC 50V) | First effective range:0.01MΩ~5MΩ ±5%rdgã€10MΩ~50MΩ ±30%rdg Second effective range: |
Accuracy(DC 125V) | First effective range:0.02MΩ~10MΩ ±5%rdg Second effective range:0.01MΩ~0.02MΩ 10MΩ~20MΩ ±10%rdg,10MΩ~100MΩ ±30%rdg |
Accuracy(DC 250V) | First effective range:0.05MΩ~20MΩ ±5%rdg Second effective range:0.02MΩ~0.05MΩ 20MΩ~50MΩ ±10%rdg,50MΩ~100MΩ ±30%rdg |
Accuracy(DC 500V) | First effective range:0.1MΩ~50MΩ ±5%rdg Second effective range:0.05MΩ~0.1MΩ 50MΩ~100MΩ ±10%rdg |
Accuracy(DC 1000V) | First effective range:2MΩ~1000MΩ ±5%rdg Second effective range:1MΩ~2MΩ 1000MΩ~2000MΩ ±10%rdg |
AC voltage (50/60Hz) | Range/Accuracy/Input impedance/Maximum input voltage AC 600V/±2.5% of full scale/Approx. 1.5M/AC 600V rms |